微孔物质BET测试数据的筛选、处理和检验
|
|
张建辉,陈小娟,余小岚,张伟庆
|
Screening, Processing and Checking of BET Test Data
|
|
Jianhui ZHANG,Xiaojuan CHEN,Xiaolan YU,Weiqing ZHANG
|
|
表2 用不同变换方法筛选处理后的BET结果对比
|
| |
|
| 序号 | 变换方法及作图选点范围 | 吸附点数 | p/p0范围 | C | Vm/(cm3∙g−1) | r | | 1 | BET变换,不筛选 | 13 | 0.005–0.3 | −169 | 109 | 0.9971 | | 2 | BET变换,经典区间 | 6 | 0.05–0.3 | −47 | 103 | 0.9966 | | 3 | BET变换,Rouquerol变换取点 | 6 | 0.005–0.03 | 3452 | 136 | 1.0000 | | 4 | Rouquerol变换倒数 | 6 | 0.005–0.03 | 3482 | 136 | $ \underline {{{0.9989}^*}} $ | | 5 | 下限小于p/p0 = 1/(1 + C1/2)上限Rouquerol变换 | 3 | 0.01–0.03 | 3434 | 136 | 1.0000 |
|
|
|