电位法基体效应在氟离子选择性电极分析中的影响
赵玲丽, 张嘉捷, 徐孝菲, 习玲玲, 委育秀, 唐云湖, 林旭锋

Matrix Effects in Potentiometric Method on Fluoride Selective Electrode Analysis
Lingli Zhao, Jiajie Zhang, Xiaofei Xu, Lingling Xi, Yuxiu Wei, Yunhu Tang, Xufeng Lin
表5 掺杂乙酸的复杂基体样品的标准曲线
乙酸含量/%回归方程浓度区间/(mol∙L−1)R2
0E = −5.61 × 10−2lgcF− + 1.94 × 10−21.00 × 10−5–1.00 × 10−20.999
5E = −5.69 × 10−2lgcF− + 2.86 × 10−21.00 × 10−5–1.00 × 10−20.999
10E = −5.91 × 10−2lgcF− + 2.69 × 10−31.00 × 10−5–1.00 × 10−20.999
20E = −5.86 × 10−2lgcF− + 3.99 × 10−31.00 × 10−5–1.00 × 10−20.999
30E = −5.90 × 10−2lgcF− + 4.60 × 10−21.00 × 10−5–1.00 × 10−20.999